4.4 digit industrial multimeter CMM-40; CAT III/1000V TRMS AC/DC

Brand: Sonel
Product code: WMGBCMM40
Price:
174,00 €excl. VAT

Industrial multimeter CMM-40 was designed for the purpose of measurements of AC/DC voltage, AC/DC current, resistance, capacitance, frequency (electrical and electronic), duty cycle, temperature and also for testing diodes and continuity.

 

The most important features of CMM-40 are:
  • autoranging/manual range selection,
  • „DATA HOLD” function, for holding measured values,
  • „REL” function, which allows you to make measurements relative to a stored reference value,
  • „MAX/MIN” function,
  • „PEAK HOLD” function,
  • memory for 2000 measurement resultS,
  • continuity test with acoustic signalling (beeper),
  • „AUTO-OFF” function,
  • display 4¾ digits (40000 max),
  • double molded, waterproof casing.

Frequency measurement (electronic)

Range Resolution Accuracy
40,000 Hz 0,001 Hz
±(0,1% m.v. + 1 digit)
400,00 Hz 0,01 Hz
4,0000 kHz 0,0001 kHz
40,000 kHz 0,001 kHz
400,00 kHz 0,01 kHz
4,0000 MHz 0,0001 MHz
40,000 MHz 0,001 MHz
100,00 MHz 0,01 MHz not speccified
  • Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and <100kHz; 5V rms min @ 20% to 80% duty cycle and > 100kHz


Frequency measurement (electrical)

Range Resolution Accuracy
40,00 Hz...10,000 kHz 0,01 Hz...0,001 kHz ± 0,5% m.v.
  • sensitivity: 1V rms

Duty cycle measurement

Range Resolution Accuracy
0,10...99,9% 0,01% ±(1,2% m.v. + 2 digits)
  • pulse width: 100μs - 100ms,
  • frequency width: 5Hz...150kHz.

Temperature measurement

Range Resolution Accuracy
-50,0°C...+1200,0°C 0,1°C ±(1% m.v. + 2,5°C)
-58,0°F...+2192,0°F 0,1°F ±(1% m.v. + 4,5°F)

DC voltage measurement

Range Resolution Accuracy
400,0 mV 0,01 mV ±(0,06% m.v. + 4 digits)
4,0000 V 0,0001 V
40,000 V 0,001 V
400,00 V 0,01 V
1000,0 V 0,1 V ±(0,1% w.m. + 5 cyfr)


AC voltage measurement (True RMS)

Range
Resolution
Accuracy
400,00 mV 0,01 mV ±(1% m.v. + 40 digits)
4,0000 V 0,0001 V ±(1% m.v. + 30 digits)
40,000 V 0,001 V
400,00 V 0,01 V
1000,0 V 0,1 V
  • fr quency range: 50...1000Hz.


AC/DC current measurement (True RMS)

Range Resolution Accuracy
400,00 μA 0,01 μA dla DC ±(1,0% m.v. + 3 digits)

dla AC ±(1,5% m.v. + 30 digits)
4 000,0 μA 0,1 μA
40,000 mA 0,001 mA
400,00 mA 0,01 mA
10,000 A 0,001 A
  • 20A: 30 sec max with reduced accuracy.

Resistance measurement

Range Resolution Accuracy
400,00 Ω 0,01 Ω ±(0,3% m.v. + 9 digits)
4,0000 kΩ 0,0001 kΩ ±(0,3% m.v. + 4 digits)
40,000 kΩ 0,001 kΩ
400,00 kΩ 0,01 kΩ
4,0000 MΩ 0,0001 MΩ
40,000 MΩ 0,001 MΩ ±(2,0% m.v. + 10 digits)


Capacitance measurement

Range Resolution Accuracy
40,000 nF 0,001 nF ±(5,0% m.v. + 7 digits)
400,00 nF 0,01 nF
4,0000 μF 0,0001 μF ±(3,0% m.v. + 5 digits)
40,000 μF 0,001 μF
400,00 μF 0,01 μF
4000,0 μF 0,1 μF ±(5,0% m.v. + 5 digits)
40,000 mF 0,001 mF

Other technical data:

  • power supply: 9V battery type 6LR61
  • diode test: I=0,9mA, U0=2,8V DC
  • continuity test: I<0,35mA, sound signal for R<35Ω
  • over range indication: 0L displayed
  • crest factor: ≤ 3 at full scale up to 500V; decreasing linearly to ≤ 1,5 at 1000V
  • PEAK: captures peaks >1ms
  • sampling rate: 2 times per second
  • input impedance: 10MΩ (V DC), 9MΩ (V AC)
  • display: 40000 counts backlit LCD with bargraph, 4¾ digits
  • store capacitance: 2000
  • dimensions: 187 x 81 x 55 mm
  • weight: 342 g
  • fuses: mA, µA range: 0,5A/1000V ceramic fast blow; A range: 10A/1000V ceramic fast blow
  • auto power OFF: 15 min.
  • accordance with following standards: EN 61010-1:2004, EN 61010-2-032

Electric security:

  • type of insulation: double, according to EN 61010-1 and IEC 61557
  • measurement category: CAT III 1000V acc. to EN 61010-1:2004
  • protection class acc. to EN 60529: IP67

Rated operational conditions:

  • operating temperature: 0...+40°C
  • storage temperature: -20...+60°C
  • operating humidity: max 80% up to 87°F (31°C); decreasing linearly to 50% at 104°F (40°C)
  • operating altitude: max 2000m

„m.v.”- measured value